In Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, the
software saves user-provided information into a comma-separated value
(CSV) file, but it does not neutralize or incorrectly neutralizes
special elements that could be interpreted as a command when the file is
opened by spreadsheet software.
References
Link | Resource |
---|---|
https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01 | Third Party Advisory US Government Resource |
https://www.philips.com/productsecurity |
Configurations
Configuration 1 (hide)
|
History
No history.
Information
Published : 2020-09-11 13:15
Updated : 2023-12-12 21:15
NVD link : CVE-2020-16214
Mitre link : CVE-2020-16214
CVE.ORG link : CVE-2020-16214
JSON object : View
Products Affected
philips
- patient_information_center_ix
CWE
CWE-1236
Improper Neutralization of Formula Elements in a CSV File