Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.
                
            References
                    | Link | Resource | 
|---|---|
| https://www.us-cert.gov/ics/advisories/icsma-20-170-04 | Third Party Advisory US Government Resource | 
Configurations
                    Configuration 1 (hide)
| AND | 
            
            
 
  | 
    
History
                    No history.
Information
                Published : 2020-06-29 14:15
Updated : 2020-07-09 17:58
NVD link : CVE-2020-12040
Mitre link : CVE-2020-12040
CVE.ORG link : CVE-2020-12040
JSON object : View
Products Affected
                baxter
- sigma_spectrum_infusion_system
 - sigma_spectrum_infusion_system_firmware
 
CWE
                
                    
                        
                        CWE-319
                        
            Cleartext Transmission of Sensitive Information